PENGARUH VARIASI VOLUME DARAH PADA TABUNG VACUTAINER K2EDTA TERHADAP NILAI MCV
Total View This Week0
Institusion
Sekolah Tinggi Ilmu Kesehatan Nasional
Author
Mela, Permata Sari
Subject
R Medicine (General) 
Datestamp
2020-12-23 06:34:21 
Abstract :
The K2EDTA vacutainer tube is best recommended by ICSH (International Council for Standardization in Hematology) and CLSI (Clinical and Laboratory Standards Institute). Comparison of blood with inappropriate anticoagulants can give inaccurate results, especially MCV values. In the hospital practice improper filling of tubes is still common. The purpose of this study was to determine the effect of variations in blood volume in the K2EDTA vacutainer tube on MCV values. This research study of literature uses descriptive analysis sourced from journals and books in the last 10 years. Based on data analysis, the examination is carried out automatically by taking blood using the K2EDTA vacutainer tube with various variations in blood volume. The study results of P value is >0,05 it showed that was no significant effect of differences in the variation volume on K2EDTA vacutainer on MCV value. After analyzing it was concluded that the variation in blood volume in the K2EDTA vacutainer tube did not significantly influence the MCV value. 